Effect the Annealing on The Growth Length Rate of Alpha Particles Tracks in CR-39 Nuclear Track Detector | ||
Journal of Education and Science | ||
Article 1, Volume 32, Issue 3, September 2023, Pages 1-8 PDF (587.83 K) | ||
Document Type: Research Paper | ||
DOI: 10.33899/edusj.2023.138444.1335 | ||
Authors | ||
Rasha Mahmood wael* 1; Mushtaq Abed Al-Jubbori2 | ||
1Department of physics, College of Education for Pure Sciences, University of Mosul, Mosul, Iraq | ||
2Department of Physics, College of Education for Pure Science, Mosul University, Mosul-Iraq | ||
Abstract | ||
The aime of this paper is to investigate the impact of annealing on track growth rate (dL/dt), which is calculated using the track length measurement method. A 600 micro meter-thick CR-39 detector, divided into four samples (1 x 1 cm2), was vertically irradiated by alpha particles with 2.6 MeV from a 241Am source which have max energy 5.48 MeV . The three samples were annealed at 100–140oC by type 20o C using oven Memmert U4079142. All samples were etched in a 6.25 N NaOH solution at 70±1 oC in a Memmert W200 water bath, the periodic etching time was 0.5 hours to 6 hours. The etched image was captured using a camera (MADC_5A) attached to a microscope (XSZ-H Series Biological Microscope). The bulk etch rate and diameters increase with increasing annealing temperatures. The new method used in this paper to calculate the experimental track length is based on the relation between track length and track diameter obtained from track-test software. | ||
Keywords | ||
Nuclear track detectors CR-39,,; ,،,؛Heting annealing,,; ,،,؛Chemical etching,,; ,،,؛Track growth rate | ||