Preparation of UO22+ as ion selective electrode depend on active substance dibenzo-18-crown-6 (DB18C6) by using PVC polymer | ||
journal of al-qadisiyah for pure science(quarterly) | ||
Article 1, Volume 12, Issue 4, December 2012, Pages 116-128 PDF (0 K) | ||
Author | ||
Abbas noor Al-Sharify | ||
Abstract | ||
A Uranyl (II) ion- selective electrode was prepared, depend on the active material Dibenzo-18-crown-6 ( DB18C6) as a sensor, deposed on a poly vinyl chloride ( PVC) polymer, by using Di-n-Octyl phenyl phthalate (DOPP) as a ,plasticizer substance in the membrane, the behavior and characteristics properties of this electrode have been studied. It respond for the concentration range (1×10-7-1×10-1) mol/L of uranyl ion with nernast linearity slope equal to 29 mV/decade of activity and correlation coefficient was about (0.9990). The selectivity coefficient K_(i,j)^potof this electrode was calculated, in the presence of some interferencing cations and anions with certain uranyl (II) ion solutions. It was found that the range of pH response is (4-9), with response time (25) sec. at (298) K, and the electrode lifetime was found (28) days. Key words: Ion Selective Electrode, UO22+ , DB18C6 . | ||
Statistics Article View: 135 PDF Download: 69 |